· Piotr Gregorczyk · Regulation · 17 min read
AI in GxP: where the rules are heading
The draft EU GMP annex on AI, the revised Annex 11, ICH E6(R3) and the AI Act point to the same expectations: a tested plan for when the model is unavailable, meaningful human oversight, change control over the model, explainability, and monitoring that works like calibration. Where each text stands at the end of October 2026, and what it means for design.
A note before you read. I’m not a regulatory, quality or legal expert. I’m a technology leader who designs and runs systems in this industry, and this is my engineer’s reading of public texts, several of which are still drafts. Treat it as a starting point for a conversation with your QA and regulatory colleagues, not as regulatory or legal advice, and check the primary sources linked below before you act on it.
The regulators are not asking for anything engineering does not already know how to do. They are asking for it to be decided, written down and tested.
In Design for the day the model is down I described six design decisions for any AI feature: what happens without the model, the fallback, containing the output, change control, explainability and monitoring, and testing the failure. None of those depend on an industry.
This post is about the industry I work in: regulated life sciences, where “regulated” means GxP. That is GMP in manufacturing and GCP in clinical research, plus FDA’s 21 CFR Part 11 for electronic records. The EU AI Act sits across all of them as horizontal law.
The interesting thing about 2025 and 2026 is not any single text. It is that the GMP drafts, the GCP guideline, the regulator workshops and the AI Act are pointing the same way. If you design to that direction now, the final texts will mostly confirm what you already built.
Where the rules stand (end of October 2026)
Several of the texts people quote on this topic are drafts, so it is worth being precise.
| Instrument | Status | What it means for design |
|---|---|---|
| EU GMP Annex 11 (2011) | In force | Applies today. Clause 16 requires business continuity for critical processes, “e.g. a manual or alternative system”. |
| Revised Annex 11 and Chapter 4 (2025 drafts) | Consultation closed 7 Oct 2025. Not adopted: the 2011 texts still apply. | Shows direction: supplier oversight, change control, disaster recovery with a defined recovery time objective (RTO). |
| New Annex 22, Artificial Intelligence (2025 draft) | Consultation closed 7 Oct 2025. Not adopted. Scope under revision. | The draft excludes generative AI and LLMs from critical GMP use. After the consultation, the drafting group has discussed allowing them under a fully documented, risk-based control strategy. An EMA expert workshop on this was held on 30 June and 1 July 2026; its report (October 2026) records no decision yet. |
| ICH E6(R3) GCP | Principles and Annex 1 in effect in the EU since 23 July 2025. Annex 2 (decentralised, pragmatic and real-world-data trials) applies from 15 January 2027. | Computerised systems in trials must be fit for purpose, for example through risk-based validation. |
| EMA guideline on computerised systems and electronic data in clinical trials (2023) | In effect | GCP counterpart to Annex 11 expectations. AI used in trials is listed in scope, with AI-specific requirements left for a possible future annex. |
| EU AI Act, as amended by the Digital Omnibus on AI | In force. Applies in stages. | General-purpose model provider obligations since Aug 2025. Transparency and enforcement since Aug 2026. High-risk rules from Dec 2027 (Annex III) and Aug 2028 (products such as medical devices). |
Two consequences follow. First, nothing in the AI annex can be cited as a requirement yet, but its drafts and the regulator discussion around them show clearly where inspectors are heading. Second, Annex 22 is a GMP text. In clinical research the same expectations arrive through GCP: fitness for purpose, risk-based validation and data integrity.

Annex 11, Annex 22, ICH E6(R3) and the AI Act come from different bodies, but their expectations for AI point the same way.
Five directions the texts agree on
Continuity is a tested, risk-based plan, not a manual path everywhere
The regulations already allow for the difference between AI that assists a task and AI that enables a task people could not do at that scale, as long as it is a decision and not an accident:
- Annex 11 (2011), clause 16 asks for continuity provisions for critical processes, “e.g. a manual or alternative system”. The time to bring them into use “should be based on risk and appropriate for a particular system and the business process it supports”, and the arrangements should be “adequately documented and tested”. The requirement is a tested plan sized to the risk, not a manual path in every case.
- The 2025 Annex 11 draft moves this into a disaster recovery plan that ensures continuity “within a defined Recovery Time Objective (RTO)”.
- EMA’s June 2026 workshop on Annex 22 raised the case directly. Regulators noted that a fully model-operated process might not have a classical fallback. The response was that this has to be addressed in the contingency strategy, and that sometimes accepting a stoppage may be the contingency.
In GxP terms, the three deciding factors from the design post become: impact on patient safety, product quality and data integrity; time-criticality (expedited safety reporting, with deadlines of 7 or 15 days, cannot sit behind a provider outage of unknown length); and detectability. A high-impact, time-critical step with no contingency needs a manual or alternative path, however rarely it is used.
A model fallback has a specific meaning here. A different model is a different system. If it was not assessed for the same intended use before the outage, failing over to it swaps an unvalidated configuration into a validated process, at exactly the moment nobody is watching closely.
Oversight has to be meaningful, and the overseer is monitored too
Industry presenters at the EMA workshop drew a useful line. Human oversight is the broad, lifecycle-wide objective: a qualified, accountable person can understand, challenge, intervene, override, reject or escalate. Human-in-the-loop is one way to implement it, where the process waits for a human action. It is one control, not a default for every use.
Regulators also asked how to monitor the human doing the oversight and how to manage automation bias. That is a notable shift. A reviewer who accepts every suggestion has always been a weak control; it is now a named inspection topic.
Attribution stays where it was. Part 11 and Annex 11 expect you to know who did what and to keep that history. Where a person decides, accept, edit or reject is recorded against a named user. Storing model text as if the accountable person had written it skips the control.
A model change is a regulated change, and the supplier does not carry your responsibility
Annex 11 already expects controlled changes. The 2025 draft says that any significant change that may affect product quality, patient safety or data integrity “should be subject to re-qualification and validation”. The Annex 22 draft adds continuous oversight of AI models: change control, performance monitoring and human review where needed.
The workshop was plain about third-party models. The responsibility stays with the regulated user even when the model is an external API. If a third-party model or API cannot provide the outputs, logs or evidence needed for control, the manufacturer cannot measure and justify the system adequately. In practice that means supplier assessment under GAMP 5 covers version retirement notice, change notification, processing location and access to evidence, and that the model, prompt, retrieval corpus and route are all configuration items.

Explainability and confidence are part of the draft Annex 22, not optional extras added after go-live.
Explainability and confidence are written into the draft
The Annex 22 draft asks that, during testing of models in critical GMP applications, systems “capture and record the features in the test data that have contributed to a particular classification or decision”. Where applicable, it names the techniques: “feature attribution (e.g. SHAP values or LIME) or visual tools like heat maps” (section 8.1).
Attribution alone is not the control. Section 8.2 asks for feature justification: a review, as part of approving the test results, that the model decides on relevant and appropriate features. A visual inspection model that rejects vials because of the lighting in one corner of the image has passed a test for the wrong reason.
Section 9 adds confidence. During testing, where applicable, the system logs a confidence score for each prediction or classification, and the model has a threshold so that it only decides when that is suitable. If confidence is very low, the draft asks you to consider flagging the outcome as “undecided” rather than guessing.
These sections were written for classifiers and predictors with defined input features. They fit generative models poorly, which is part of why the draft excluded LLMs from critical use. At the June 2026 workshop, an industry presentation listed what the practical equivalents look like for generative features: answers bound to retrieved sources with citations, rule-based limits and hard constraints on the output, uncertainty or self-consistency scores, and confidence gates. If the scope is widened to allow LLMs under a control strategy, I expect these to be the controls that strategy is judged on.
Monitoring is treated like calibration
Section 10 of the Annex 22 draft is about everything after acceptance testing:
- Change control (10.1). The model, the system it runs in and the process it supports are under change control before deployment. Any change is assessed for retesting, and a decision not to retest “should be fully justified”.
- Configuration control (10.2). The tested configuration is locked, with “effective measures” to detect unauthorised change.
- System performance monitoring (10.3). The model’s performance, as defined by its metrics, is monitored regularly in operation.
- Input sample space monitoring (10.4). Inputs are checked regularly to confirm they are still within the sample space and intended use the model was tested on, with defined metrics for drift.
- Human review records (10.5). Where the model gives input to a decision made by a person, and the testing effort was reduced because of that, records are kept of the review. Depending on criticality and the level of testing, that may mean reviewing every output.
An industry speaker at the workshop made the connection explicit. Drift is not new to GMP. Measurement systems drift too, and calibration intervals and lifecycle review already exist to confirm that a system remains fit for use. Presenters also listed CUSUM and EWMA charts, confidence gates and fallback to validated states. The golden set, input drift checks and human decision rates from the design post are how you produce that evidence for a hosted model.

The AI Act is risk-based: the obligations depend on how a system is used and classified, not on the technology alone.
Where the AI Act fits
The AI Act is the instrument that reaches almost every model you will use. It is horizontal law about AI systems and models, not a quality-system framework. Here is how it maps onto a regulated workflow. This is a design view, not legal advice.
Your vendor carries the model obligations. Providers of general-purpose AI models, which covers essentially every frontier LLM, have had obligations since 2 August 2025: technical documentation, information for downstream providers, a copyright policy and a public summary of training content. Models already on the market before that date have until 2 August 2027. The AI Office’s enforcement powers started on 2 August 2026. Use this in supplier assessment. Ask the model provider for the documentation they are required to make available to downstream providers.
Know your own role. If you use a vendor’s AI product as supplied, you are a deployer. If you build a feature on top of a model and put it into service under your own name, including for internal use, you are likely also the provider of that AI system. The role decides which obligations land on you.
Most GxP assistants are not high-risk under the Act. High-risk status comes from Annex III use cases (for example employment, education, access to essential services including some health-related ones, and emergency triage) or from AI that is part of a regulated product requiring third-party conformity assessment, such as many medical devices. A drafting assistant in document control is neither. AI in a medical device is the main exception, with rules applying from 2 August 2028 after the Digital Omnibus.
The high-risk controls are still the right reference. Human oversight must allow a person to “interrupt the system through a ‘stop’ button or a similar procedure that allows the system to come to a halt in a safe state” (Article 14(4)(e)). Robustness “may be achieved through technical redundancy solutions, which may include backup or fail-safe plans” (Article 15(4)). That is the per-feature kill switch and the designed fallback, in legal language.
Two obligations apply now regardless of risk class:
- AI literacy (Article 4). Providers and deployers must take measures to support AI literacy among staff using AI systems. The Omnibus softened the wording: no guaranteed level per individual. In GxP terms this is training, which you need anyway for meaningful oversight.
- Transparency (Article 50), since 2 August 2026. People interacting directly with an AI system must be told so, unless it is obvious, and generated content must be marked in a machine-readable, detectable way (provider obligations; systems already on the market before 2 August 2026 have until 2 December 2026 for the marking). Deployers must disclose AI-generated text published to inform the public on matters of public interest, unless a person has reviewed it and someone holds editorial responsibility. Patient-facing material, such as published lay summaries, is where this is most likely to come up. Human review with named responsibility is the same control GxP already asks for.
The instruments are separate. Compliance with one does not demonstrate compliance with the other, and a delay in one does not move the other.
What if your system is not under Annex 11 today?
Many AI features are not in Annex 11 scope today. Annex 11 applies to computerised systems used in GMP-regulated activities. A drafting assistant for internal reports, a research tool before any GMP stage or a support chatbot may sit entirely outside it. Medical device software follows its own framework. Clinical trial systems fall under GCP, not GMP.
That is a real distinction, and it should change how much you do. It does not mean the principles stop applying. Here is why.
Scope follows use, not the system
A tool is out of scope until its output gets pasted into a deviation report, a batch record or a trial document. From that moment the record is regulated, whatever the tool’s classification says. Out-of-scope AI features need a defined boundary: what their output may and may not be used for, and ideally a technical control that keeps it out of regulated records, or at least labels it on the way in. The 2025 Annex 11 draft states the underlying rule plainly: when a computerised system replaces a manual operation, “there should be no increase in the overall risk of the process.”
Other frameworks ask for the same controls
Leaving GMP does not leave regulation:
- GCP. ICH E6(R3) requires computerised systems in trials to be fit for purpose, for example through risk-based validation.
- GDPR. Where personal data is processed, Article 32 asks for the “ability to ensure the ongoing confidentiality, integrity, availability and resilience of processing systems and services” and to “restore the availability and access to personal data in a timely manner”. An AI feature processing patient or participant data is a processing system.
- NIS2. Annex I lists the health sector, including entities doing research and development of medicinal products and entities manufacturing pharmaceutical products. Whether a given company is covered depends on size thresholds and national transposition. For entities in scope, Article 21 lists business continuity (such as backup management and disaster recovery) and supply chain security among the required measures. An AI provider is part of the supply chain.
- AI Act. AI literacy and transparency apply regardless of GxP scope.

Audit trails and records cannot be reconstructed later, so logging has to start on the first day of use.
Scope changes
A feature that starts as an internal helper can become part of a regulated process through a new use case, a customer requirement or a sponsor audit. Pinned versions, a change log and an audit trail are cheap to build in from the start and expensive to reconstruct later. You cannot reconstruct last year’s audit trail.
Failures do not check the scope
A provider outage or a silent model change hurts an unregulated process too. It is just that nobody inspects the damage.
So scale the controls down, not off. This matches the risk-based thinking in GAMP 5. A reasonable minimum for an AI feature outside Annex 11:
| Control | Outside Annex 11 | Inside Annex 11 |
|---|---|---|
| Decision on what happens without the model | Yes, brief | Yes, documented in the risk assessment and tested |
| Kill switch with a designed off state | Yes | Yes |
| Pinned model and prompt versions | Yes, with a change log | Yes, under formal change control |
| Golden set re-run after changes | Yes | Yes, plus scheduled monitoring with control limits |
| AI-generated label and sources | Yes | Yes |
| Boundary control so output does not enter regulated records unlabelled | Yes | Not applicable: the records are already regulated |
| Attributable human decision per output | Where a decision is made | Yes |
| Pre-qualified fallback model | Optional | Required if any model fallback is used |
| Formal validation and periodic review | No | Yes |
The left column takes days to build, not months. It is also most of what you need if the feature later moves into the right column.
Checklist for regulated teams
On top of the general checklist in the design post:
- Is the continuity decision for each AI feature in the risk assessment, with a stated recovery time and a tested plan, as Annex 11 clause 16 expects?
- Is any model fallback pre-qualified for the same intended use, and named in the risk assessment?
- Is human oversight defined per use case, and is the overseer monitored for automation bias?
- Do supplier agreements give you version retirement notice, change notification and the evidence you need for control?
- For predictive models in GMP use, were feature attributions recorded and justified during testing (Annex 22 draft, section 8)?
- Is there a confidence threshold with an “undecided” outcome (section 9)?
- Is operational monitoring set up like calibration: defined metrics, control limits and a defined action (section 10)?
- For features outside Annex 11, is there a boundary that keeps their output out of regulated records, or labels it on the way in?
- Is your role under the AI Act (deployer, provider or both) known, and are AI literacy and transparency covered?
- Has someone re-checked the status of Annex 22, the revised Annex 11 and Chapter 4 this quarter?
Design to the direction, not the deadline
It is tempting to wait for the final texts. I think that is the wrong call. The drafts may change in scope and wording, but the direction is consistent across GMP, GCP and the AI Act: decide what happens without the model, keep a qualified person meaningfully in control, treat the model as a controlled configuration item, explain what it decided, and keep checking that it still behaves as approved.
None of that is new to a quality system. It is the discipline we already apply to instruments, suppliers and computerised systems, applied to a component that happens to be a model.
Sources
- European Commission, EudraLex Volume 4: current GMP annexes, including Annex 11 (2011 revision).
- European Commission, EU GMP Annex 11: Computerised Systems (2011), clause 16 Business Continuity.
- European Commission, Stakeholders’ consultation on Chapter 4, Annex 11 and new Annex 22 (7 July to 7 October 2025), with links to all three drafts.
- European Commission, Annex 22: Artificial Intelligence, consultation draft: scope (section 1), explainability with SHAP and LIME (section 8), confidence (section 9), operation and monitoring (section 10).
- PIC/S, Joint stakeholders consultation on Chapter 4, Annex 11 and Annex 22 (7 July 2025).
- EMA, Annual report of the GMDP Inspectors Working Group 2025: drafting of both annexes continues after the consultation.
- EMA, Multistakeholder workshop on AI guidance development (Annex 22), 30 June and 1 July 2026, and its report (EMA/156789/2026, published 2 October 2026).
- EMA, ICH E6(R3) Guideline for good clinical practice, principles and Annex 1, in effect from 23 July 2025; Annex 2, in effect from 15 January 2027. Overview: EMA, ICH E6 Good clinical practice.
- EMA, Guideline on computerised systems and electronic data in clinical trials (2023).
- European Commission, AI Act: regulatory framework and timeline, and the Digital Omnibus on AI, Regulation (EU) 2026/1744 (in force 27 July 2026).
- AI Act Service Desk, Implementation timeline, and consolidated text of Article 4, Article 14, Article 15 and Article 50.
- Regulation (EU) 2016/679 (GDPR), Article 32(1)(b) and (c).
- Directive (EU) 2022/2555 (NIS2), Article 21(2)(c) and (d), and Annex I (health sector).
- European Commission, Transparency obligations under Article 50: FAQ.
Author
Piotr Gregorczyk
I'm a strategic technology executive with 15+ years in CTO and CIO roles, after a foundational decade at Accenture.
An engineer at heart and a successful manager for years, I hold an MSc in Computer Engineering from Warsaw University of Technology and a recent Executive MBA.
I turn technology modernization into real business value, with experience across clinical research, insurance, banking, real estate, software and consulting.
The opinions on this site are my own and do not represent any company I work or have worked for.